Detector Type
Thermal Conductivity Detector (TCD)
- Detection Object: ppm-% impurities in gas
- Detection Limit: 10ppm
- Carrier Gas Type: N2/Ar/He
- Carrier Gas Flow: 100ml/min
Hydrogen Flame Ionization Detector (FID)
- Detection Object: ppm-% hydrocarbon impurities in gas
- Detection Limit: 0.03ppm
- Carrier Gas Type: N2/Ar
- Carrier Gas Flow: 100ml/min
Pulse Discharge Ammonic Ionization Detector (PDHID)
- Detection Object: ppb level trace impurities in high purity gas
- Detection Limit: 5ppb
- Carrier Gas Type: He
- Carrier Gas Flow: 150ml/min
Plasma Detector (PED)
- Detection Object: ppb level trace impurities in high purity gas
- Detection Limit: 0.5ppb
- Carrier Gas Type: N2/Ar/He/CO2
- Carrier Gas Flow: 150ml/min
Gas Path Control
Sample Gas Flow: Positive Pressure: 50ml/min or negative pressure sampling
Sample Gas pressure: -0.1MPa ~ 0.1MPa
Sample Gas Interface: 1/4
Carrier Gas Pressure: 5 ~ 30 PSIG
Flow Control: Electronic Monitoring
Electrical Appliance
Power Supply: AC220V, 50/60Hz
Consumption: 400W
Communication Interface: RS232/485, TCP
Protocol: Modbus-RTU, Modbus-TCP, 4-20mA
Shape
Dimension: Standard 5U Rack-mount Chassis
Temperature Control Parameters
Temperature Control Area: 8 ways
Temperature Control Range: Above room temperature 4°C ~ 450°C, increment: 1°C, Accuracy: ±0.01°C
Program Heating Steps: 20 steps